Tuesday, 20 October 2009 08:41 |
|
Soft X-Ray Microscopy
Scientific disciplines: Magnetism, spin dynamics, x-ray optics, environmental science, materials science
| GENERAL BEAMLINE INFORMATION |
|
Operational
|
Yes
|
| Proposal cycle |
Proposals for General Sciences Beamlines (6-month cycle) |
|
Source characteristics
|
Bend magnet
|
|
Energy range
|
500-1300 eV
|
|
Monochromator
|
Zone-plate linear
|
|
Measured flux (1.9 GeV, 400 mA)
|
Images with 1000 x 1000 pixels, 1000 photons/pixel recorded in 3 s at 517 eV with 0.2% BW
|
|
Resolving power (E/ΔE)
|
500-700
|
|
Endstations
|
X-ray microscope (XM-1)
|
|
Characteristics
|
Full-field soft x-ray transmission microscope
|
|
Spatial resolution
|
Typical 25 nm (best value 12 nm)
|
|
Detectors
|
Back-thinned 2048 x 2048-pixel CCD camera
|
| Equipment |
Various fast electronic pulsers for spin dynamics, fast sampling oscilloscope, polarization selecting aperture
|
|
Field of view
|
15 µm single field; virtually unlimited larger areas can be tiled together like a mosaic
|
|
Sample format
|
Thin films (OD=1, for specific sample calculation see http://www.cxro.lbl.gov/optical_constants/) deposited typically on Si3N4 membrane (similar to TEM substrates) or other x-ray-transparent substrates
|
|
Sample preparation
|
Sample dependent
|
|
Sample environment
|
External magnetic fields up to 5 kOe in beam direction and 2 kOe along the sample plane; tilted sample holder for in-plane magnetized samples; He at atmospheric pressure, wet or dry
|
|
Special notes
|
Mutual indexing system with visible-light microscopy provided to position and focus sample
|
| Scientific disciplines |
Magnetism, spin dynamics, x-ray optics, environmental science, materials science |
|
Scientific applications
|
Transmission imaging x-ray microscopy for magnetic materials (XMCD contrast), materials sciences, environmental sciences.
|
|
Experimental Techniques
|
Microscopy, Magnetic Dichroism, Timing, Ultrafast, Tomography.
|
|
Local contact / Spokesperson
|
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
Center for X-Ray Optics, Berkeley Lab Phone: (510) 486-7052 Fax: (510) 486-4550
|
|
Beamline phone number
|
(510) 495-2061
|
| Website |
CXRO: http://www.cxro.lbl.gov/BL612  |
|