Scientific discipline: Structural biology
| GENERAL BEAMLINE INFORMATION |
|
Operational
|
Yes
|
| Proposal cycle |
Proposals for Structural Biology Beamlines (2-month cycle) |
|
Source characteristics
|
Superbend
|
|
Energy range
|
5-17 keV (1% max flux)
|
| Frequency range |
2.5-0.73 Angstrom wavelength |
| Beam size |
100 µm round beam default 10 µm and 30 µm collimators available for small samples with flux reduced to 1% and 12%, respectively. 120 µm x 120 µm Gaussian uncollimated beam shape at sample.
|
| Scientific discipline |
Structural biology |
|
Monochromator #1
|
Si(111) Double crystal
|
|
Calculated flux (1.9 GeV, 400 mA)
|
2.5 x 1011 photons/sec at 11 keV
|
|
Resolving power (E/ΔE)
|
7000
|
|
Monochromator #2
|
0.6% bandpass multilayers
|
|
Calculated flux (1.9 GeV, 400 mA)
|
4 x 1013 photons/s with a convergence angle of 3 mrad
|
|
Resolving power (E/ΔE)
|
100
|
|
Special notes
|
Computers for data processing and analysis are available.
|
|
Local contacts
|
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
Physical Biosciences Division, Berkeley Lab Phone: (510) 486-5378 Fax: (510) 486-5298
|
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
Physical Biosciences Division, Berkeley Lab Phone: (510) 495-2697 Fax: (510) 486-5298
|
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
Physical Biosciences Division, Berkeley Lab Phone: (510) 486-5378 Fax: (510) 486-5298
|
|
Spokesperson
|
This e-mail address is being protected from spambots. You need JavaScript enabled to view it
The Scripps Research Institute Phone: (858) 784-8119 Fax: (858) 784-2289
|
| Beamline phone number |
(510) 495-2134 |
| Website |
SIBYLS: http://bl1231.als.lbl.gov/  |
| ENDSTATION INFORMATION |
|
Endstation name
|
Macromolecular crystallography
|
|
Detector
|
3 x 3 array CCD, ADSC Q315
|
| Spatial resolution |
~1 µm pixel video microscope |
|
Calculated spot size at sample (FWHM)
|
10-100 µm (collimated)
|
|
Sample format
|
Single-crystal biomolecules
|
|
Sample preparation
|
Support labs available (Bldg. 6, Rm. 2105)
|
|
Sample environment
|
100-300K
|
|
Scientific applications
|
Macromolecular crystallography, multiple-wavelength anomalous diffraction (MAD)
|
| ENDSTATION INFORMATION |
|
Endstation name
|
Small-angle X-ray scattering
|
|
Detector
|
Mar 165 CCD
|
|
Calculated spot size at sample (FWHM)
|
~5.0 x 0.5 mm
|
|
Sample format
|
~15 µL solutions of macromolecules and buffer blanks
|
|
Sample preparation
|
Support labs available (Bldg. 6, Rm. 2105)
|
|
Sample environment
|
80 - 4C, anaerobic
|
|
Scientific applications
|
High-throughput solution structural analyses by SAXS. Solution structure modeling, molecular envelope shape reconstruction, identification and characterization of oligomeric states and macromolecular complex assembly.
|
More information about Beamline 12.3.1 is available on the SIBYLS Web page .
|